Flow characteristics of cerebrospinal fluid shunt tubing

14

Citations

12

References

2012

Year

Abstract

There appears to be some increase in resistance of CSF shunt catheters as they age, altering flow dynamics. In addition, the use of straight connectors within a CSF shunt system increases the resistance to flow of artificial CSF within the shunt system. The increase in resistance appears to be related to the duration of implantation and the length of the catheter and inversely related to the diameter of the catheter. This increase in resistance may be related to sterile shunt malfunction. The addition of straight connectors is associated with a significant increase in resistance in comparison with catheters without connectors (p = 0.005).