Artificial Organs · 2007 · 18 citations · 23 references
MedicineBiomedical EngineeringPhysiologyHemodynamic ManagementResuscitationVascular SurgeryExtracorporeal CirculationExtracorporeal Life SupportFluid ManagementFluid ResuscitationFluid TherapyHemodynamic ResuscitationDynamic Flow MeasurementIn Vivo StudiesPump Speed ControlVenous Flow DynamicsVenous Hemodynamic AssessmentVenous PhysiologyVenous ReturnVenous Return Assessment
The objective of this study was to investigate venous collapse (VC) related to venous drainage during the use of an extracorporeal life support circuit. A mock circulation was built containing a centrifugal pump and a collapsible vena cava model to simulate VC under controlled conditions. Animal experiments were performed for in vivo verification. Changing pump speed had a different impact on flow during a collapsed and a distended caval vein in both models. Flow measurement in combination with pump speed interventions allows for the detection and quantitative assessment of the degree of VC. Additionally, it was verified that a quick reversal of a VC situation could be achieved by a two-step pump speed intervention, which also proved to be more effective than a straightforward decrease in pump speed.
23
Robert H. Bartlett, Dietrich W. Roloff, Joseph R. Custer et al. · JAMA · 2000 · 353 citations
Medicine, Cardiology, Pediatrics +14
Development of a Suction Detection System for Axial Blood Pumps
Michael Vollkron, Heinrich Schima, Leopold Huber et al. · Artificial Organs · 2004 · 108 citations