IEEE Microwave and Wireless Components Letters · 2012 · 26 citations · 11 references
EngineeringBiomedical EngineeringElectrical EngineeringBiosensorsElectroanalytical MethodsElectromagneticsBiomedical InstrumentationBiopotential MeasurementElectrical CharacterizationElectrical MeasurementElectro-optic EffectsUltrafast OpticsElectric Field SensingElectroporationMeasurement Uncertainty In TemperatureTemperature-dependent Electrical Measurements
High intensity nanosecond pulsed electric fields and temperature were simultaneously measured using a unique electro-optic (EO) probe. The measurements were performed in an electroporation cuvette with 4 mm electrode gap and filled with a buffered salt solution. High voltage generators delivering 2.6 and 10 ns duration pulses with different pulses shape and intensity were investigated. The EO probe linearity was characterized up to 2 MV/m. The temperature measurement uncertainty was found to be less than 22 mK. Excellent measurement abilities were achieved with this EO probe showing its suitability for bioelectromagnetic experiments and particularly for wideband high intensity field applications.
11
Ultrashort electrical pulses open a new gateway into biological cells
Karl H. Schoenbach, R. P. Joshi, Juergen F. Kolb et al. · Proceedings of the IEEE · 2004 · 393 citations
Calcium bursts induced by nanosecond electric pulses
P. Thomas Vernier, Yinghua Sun, Laura Marcu et al. · Biochemical and Biophysical Research Communications · 2003 · 392 citations
Life Sciences, Biophysics, Physics +10