Sensors · 2004 · 67 citations · 6 references
EngineeringBioengineeringMaterials ScienceAnalytical ChemistryBiosensorsSurface ScienceBiological SensorsBiosensors For Immune DetectionElectrochemical BiosensorsElectrochemical Impedance SpectroscopyEnzyme-linked Immunosorbent AssayNanomaterial-based BiosensorsChemical Surface FunctionalizationGold ElectrodesImmunosensorsQuartz Crystal MicrobalanceQuartz Crystal Microbalance BiosensorsQuartz Crystal Microbalance ImmunosensorsSurface Functionalization For BiosensingVoltammetry
In order to develop a robust biosensor based on quartz crystal microbalance technique for antigen detection, a control of the steps of the surface functionalization has been performed by impedance spectroscopy. The gold electrode is functionalized with the self-assembled monolayer technique. The high insulating properties of the acidic thiol monolayer has been characterized with cyclic voltammetry and impedance spectroscopy. The modified surface is activated with N-hydroxysuccinimide(NHS) and 1-(3-(dimethylamino)propyl)-3-ethylcarbodimide hydrochloride(EDC) cross-linker for antibody coupling. The non-specific sites are blocked with bovin serum albumine molecules. Different concentrations of antigen can be detected with a good reversibility in real time with the quartz crystal microbalance.
6
Heiko Hillebrandt, A. Abdelghani, C. Abdelghani-Jacquin et al. · Applied Physics A · 2001 · 41 citations