Concepedia

Publication | Closed Access

Investigation of American Association of Physicists in Medicine Report 204 Size-specific Dose Estimates for Pediatric CT Implementation

132

Citations

7

References

2012

Year

Abstract

The combination of AP and lateral measurements should be used to determine SSDE correction factors when possible. For pediatric patients, CTDI(vol) calculated with a 32-cm phantom requires SSDE conversion to more accurately estimate patient dose; CTDI(vol) calculated with a 16-cm phantom for pediatric patients weighing 36 kg or less does not require SSDE conversion.

References

YearCitations