A high resolution imaging susceptometer

Yukun Ma, Ian M. Thomas, A. Lauder, John P. Wikswo

IEEE Transactions on Applied Superconductivity · 1993 · 19 citations · 5 references

Concepts

Abstract

A high-resolution, MicroSQUID (superconducting quantum interference device)-based susceptometer has been constructed for imaging the susceptibility distribution in diamagnetic and paramagnetic objects. The maximum available applied field, produced by the Helmholtz pair, is 300 mu T, which provides an adequate signal-to-noise ratio for common susceptible materials. The calculated variation in the applied field within 20 mm of the center is less than 0.1% of the field at the center. In a 100-Hz bandwidth, the minimum detectable susceptibility-induced field change, (limited by SQUID noise) is 3*10/sup -9/ of the applied field. The sensitivity of the system may be improved by increasing the applied field, for instance by incorporating a superconducting magnet into the magnetometer dewar, and by using lower-noise SQUIDs.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">&gt;</ETX>