Electrical Proarrhythmia: Induction of Inappropriate Atrial Therapies due to Far‐Field R Wave Oversensing in a New Dual Chamber Defibrillator

Christian Wolpert, Werner Jung, Christopher Scholl, S. Spehl, JOACHIM CYRAN, Berndt Lüderitz

Journal of Cardiovascular Electrophysiology · 1998 · 39 citations · 5 references

Concepts

Abstract

This case report describes delivery of atrial therapies during a sinus tachycardia in a new dual chamber implantable cardioverter defibrillator inappropriately caused by far-field oversensing of ventricular beats in the atrial channel. Upon classification of the PR interval pattern, the rate criterion for an atrial tachycardia was fulfilled, and the device initiated high-frequency burst pacing as the first stage of programmed tiered atrial therapies. Atrial fibrillation subsequently was induced by high-frequency burst pacing, and eventually a programmed 10-J shock was delivered for successful termination of atrial fibrillation. The phenomenon of far-field oversensing of ventricular beats could be repeatedly observed during exercise testing and abolished by decreasing the atrial sensitivity.