MedicineOrthopaedic SurgeryOrthopedic SurgeryBiomechanicsJoint Replacement SurgeryOrthopedicsArthroplasty BiomechanicsBone BiomechanicsBone MechanicsJoint BiomechanicsKnee BiomechanicsProsthetic BiomechanicsBiomechanics Of Fixed ProsthesesJoint Replacement MechanicsKnee Implant BiomechanicsSurface AbrasionTibial Component DesignTotal Knee Arthroplasty BiomechanicsWear
Experiments were carried out to determine the optimum conformity between the femoral and tibial condyles in condylar replacement knee prostheses. Wear tests and observations from removed prostheses indicated that both high and low conformity produced characteristic abrasion and fatigue. Partly conforming condyles provided stability under load-bearing but allowed laxity to occur. Fixation to resist the various forces on the tibial components was enhanced by a short central intramedullary peg. Partial conformity is proposed as the optimum configuration between femoral and tibial components.
6
Wear and conformity in total knee replacement
PETER s. TRENT, P. S. Walker · Wear · 1976 · 18 citations