Química Nova · 2010 · 42 citations · 1 references
Natural SciencesChemistryToxicologyAnalytical ChemistryMetrologyUncertainty QuantificationAnalytical CalibrationAnalytical MeasurementsCalibrationInstrumental AnalysisIon ChromatographyMeasurement UncertaintyExternal CalibrationExternal Standard CalibrationExternal StandardizationInternal CalibrationInternal Standard CalibrationInternal StandardizationInternal StandardsStandardization
Traditionally, in the cigarettes industry, the determination of ammonium ion in the mainstream smoke is performed by ion chromatography. This work studies this determination and compares the results of this technique with the use of external and internal standard calibration. A reference cigarette sample presented measurement uncertainty of 2.0 μg/cigarette and 1.5 μg/cigarette, with external and internal standard, respectively. It is observed that the greatest source of uncertainty is the bias correction factor and that it is even more significant when using external standard, confirming thus the importance of internal standardization for this correction.