PharmacyChemical EngineeringChemistryAnalytical ChemistryElectroanalytical MethodsElectrochemistryElectrochemical DetectionElectrochemical SensorsIon-selective ElectrodesMembrane CharacterizationPolymer MembranesHeavy Metal Ion SensorsIon-selective MembranesLead PollutionPolymeric Membrane ElectrodesResponse Time
A poly(vinyl chloride)(PVC)-based membrane of 15-crown-5 exhibits a good response for lead(II) ions over a wide concentration range. The response time of the sensor is 30 s and the membrane can be used for more than four months without observing any divergence. The selectivity of the sensor is comparable with those reported for other such electrodes. It was possible to determine lead in polluted waters using this electrode assembly.
9
Lead(II) ion-selective electrodes based on crown ethers