Publication | Closed Access
How to Appropriately Calculate Effective Dose for CT Using Either Size-Specific Dose Estimates or Dose-Length Product
53
Citations
19
References
2015
Year
Current use of DLP to calculate effective dose was shown to be deficient because of the outdated means by which the k-coefficients were derived. This study shows a means to calculate EICRP using patient size-specific dose estimate and how to appropriately correct EDLP.