The conductivity of the human skull: results of in vivo and in vitro measurements
465
Citations
23
References
2000
Year
AnatomyBiomedical EngineeringSocial SciencesStimulation DeviceCerebrospinal FluidIntracranial PressureNeurologyDance ImagesFresh SkullSkull ConductivityNeurostimulationNervous SystemCerebral Blood FlowHuman SkullNeurophysiologyNeuroanatomyPhysiologyVitro MeasurementsBrain ElectrophysiologyElectrophysiologyNeuroscienceCentral Nervous SystemMedicine
The conductivity of the human skull was measured both in vitro and in vivo. The in vitro measurement was performed on a sample of fresh skull placed within a saline environment. For the in vivo measurement a small current was passed through the head by means of two electrodes placed on the scalp. The potential distribution thus generated on the scalp was measured in two subjects for two locations of the current injecting electrodes. Both methods revealed a skull conductivity of about 0.015 (1//spl Omega/)/m. For the conductivities of the brain, the skull and the scalp a ratio of 1:1/15:1 was found. This is consistent with some of the reports on conductivities found in the literature, but differs considerably from the ratio 1:1/80:1 commonly used in neural source localization. An explanation is provided for this discrepancy, indicating that the correct ratio is 1:1/15:1.