Biomechanical Evaluation of Classic Solid and Novel All‐Soft Suture Anchors for Glenoid Labral Repair

Augustus D. Mazzocca, David M. Chowaniec, Mark P. Cote, Joel Fierra, John M. Apostolakos, Michael Nowak, Robert A. Arciero, Knut Beitzel

Arthroscopy The Journal of Arthroscopic and Related Surgery · 2012 · 130 citations · 14 references

Concepts

Abstract

PURPOSE: To evaluate the biomechanical performance of an all-soft suture anchor (JuggerKnot; Biomet, Warsaw, IN) in comparison with a classic solid suture anchor (2.4-mm biocomposite SutureTak; Arthrex, Naples, FL) in an in vitro labral repair model. METHODS: We dissected 12 cadaveric shoulders (mean age, 61 ± 9.4 years), leaving the labrum intact, and bone mineral density was obtained (mean, 0.375 ± 0.06 g/cm(3)). Simulated labral tears were made at the anteroinferior and posteroinferior edges of the labrum. Repairs used 2 all-soft suture anchors (JuggerKnot) or 2 solid anchors with free, high-strength No. 2 suture (FiberWire; Arthrex) spanning the operative construct to load the repair. Differential variable reluctance transducers were used to measure labral displacement for each specimen. The testing protocol consisted of a preconditioning phase at 10 N for 10 cycles (1 Hz) and then a final load-to-failure testing at a rate of 3 mm/min. Labral displacement of 2 mm was determined as the primary outcome. RESULTS: There was no statistical difference (P = .22) in ultimate load to failure and displacement at ultimate failure (anchor pullout) between the all-soft JuggerKnot (146.0 ± 43.0 N and 19.8 ± 5.4 mm, respectively) and the solid SutureTak (171.9 ± 52.6 N and 22.3 ± 6.8 mm, respectively). The solid anchor had a significantly higher ultimate load at 2 mm of labral displacement than the all-soft suture anchor (84.1 ± 19.0 N and 39.2 ± 10.6 N, respectively; P < .001). CONCLUSIONS: Whereas both the solid SutureTak and the all-soft JuggerKnot displayed similar results on ultimate load-to-failure testing, the solid anchor required significantly greater load for 2 mm of labral displacement than the all-soft anchor. CLINICAL RELEVANCE: The all-soft anchor (JuggerKnot) is similar in biomechanical performance to the classic solid anchor (SutureTak) with the exception of load at 2 mm of labral displacement, suggesting micromotion of the device.

References

14