Magnetic Resonance in Medicine · 1994 · 12 citations · 20 references
MedicineBiomedical EngineeringCardiologyDiagnostic ImagingDiagnostic RadiologyMedical PhysicsCardiac ImagingCardiovascular ImagingMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsMagnetic Resonance Signal FormationMagnetic Resonance SpectroscopyAdiabatic PulsesMagnetic Resonance Imaging AcquisitionMagnetic Resonance Spectroscopic ImagingMagnetization Preparation PulsesMyocardial Ischemia MarkersMyocardial Tissue CharacterizationPhantom Validation
A technique is presented for MRI tagging in the presence of inhomogeneous B1 fields. A rectangular tagging grid is produced with B1-insensitive adiabatic pulses in a magnetization preparation period that precedes image acquisition. Phantom results demonstrate that the method is well-suited to surface coil experiments. The technique is applied to a canine model of myocardial ischemia to track the spatially dependent wall motion of the left ventricle during the cardiac cycle. Transmural 31P spectra are acquired from the same double-tuned surface coil, with tagging and spectroscopy performed for the first time, during normal, ischemic, and recovery conditions for the same animal.
20
Human heart: tagging with MR imaging--a method for noninvasive assessment of myocardial motion.
Elias A. Zerhouni, Dave Parish, Walter J. Rogers et al. · Radiology · 1988 · 1.4K citations
Medicine, Cardiology, Radiology +15
MR imaging of motion with spatial modulation of magnetization.
Leon Axel, L Dougherty · Radiology · 1989 · 1.2K citations