Clinical Evaluation of the ZstatFlu-II Test: a Chemiluminescent Rapid Diagnostic Test for Influenza Virus

Marilyn S. Hamilton, David Abel, Yolanda Ballam, Mary K. Otto, Angela F. Nickell, Lisa M. Pence, James R. Appleman, Craig Shimasaki, Komandoor E. Achyuthan

Journal of Clinical Microbiology · 2002 · 35 citations · 18 references

DOIFull text

Open access

Concepts

Abstract

Exploiting the high sensitivity of the chemiluminescence phenomenon, an accurate and sensitive point-of-care test, called the ZstatFlu-II test (ZymeTx, Inc., Oklahoma City, Okla.), was developed to detect influenza virus infections. The ZstatFlu-II test takes 20 min and requires approximately 2 min of "hands-on" time for operational steps. The ZstatFlu-II test does not distinguish between infections with influenza virus types A and B. ZstatFlu-II test results are printed on Polaroid High-Speed Detector Film, allowing test results to be archived. A prototype version of the ZstatFlu-II test was evaluated during the 2000-to-2001 flu season with 300 nasal aspirate specimens from children at a pediatric hospital. Compared to culture, the ZstatFlu-II test had 88% sensitivity and 92% specificity. The Directigen test had a sensitivity of 75% and a specificity of 93%. The sensitivity of the ZstatFlu-II test was significantly higher than that of the Directigen test (P < 0.0574).