Electroanalysis · 2010 · 27 citations · 15 references
Natural SciencesChemical EngineeringChemistryAnalytical ChemistryElectroanalytical MethodsElectrochemistryElectroanalytical ChemistryElectrochemical DetectionElectrode MaterialsElectrolyte SolutionsReference ElectrodesWorking ElectrodesAdsorptive Stripping VoltammetryBismuth Film ElectrodesComplexing AgentsElectrochemical MeasurementsNickel DeterminationReference Electrode CalibrationReference Electrode DesignVoltammetric Trace Analysis
Abstract For the first time an in situ plated bismuth film electrode is applied to adsorptive stripping voltammetry at mild alkaline conditions, to pH of 10. As an example of the application of such an electrode a procedure for trace Ni(II) determination by adsorptive stripping voltammetry is elaborated. For stabilization of Bi(III), tartrate as a strong complexing agent for Bi(III) was added to an ammonia buffer used as a supporting electrolyte. The appropriate concentration of Bi(III) was added to the electrolyte in the form of its complex with tartrate. The calibration graph for Ni(II) for an accumulation time of 120 s is linear from 5×10 −9 to 5×10 −8 mol L −1 . The proposed procedure was applied for Ni(II) determination in certified water reference material.
15
Allen J. Bard, James N. Butler · Journal of The Electrochemical Society · 1967 · 528 citations
Manolis Morfobos, Anastasios Economou, Anastasios Voulgaropoulos · Analytica Chimica Acta · 2004 · 128 citations