Characterisation and On-Site Monitoring of Odorous Organic Compounds in the Environment of a Landfill Site

Arnaud Termonia, M. Termonia

International Journal of Environmental & Analytical Chemistry · 1999 · 25 citations · 2 references

Concepts

Abstract

Abstract This study presents an analytical procedure developed in order to monitor specific volatile organic compounds (VOC) causing olfactive nuisances in the environment of a landfill site. l-Methyl-4-iso-propenyl-l-cyclohexene (limonene) and l-methyl-4-iso-propyl-benzene (p-cymene) detected by GC-MS in the landfill gas emissions were considered as tracers of the malodour. An on-line analytical system including a gas chromatograph coupled with a photoionisation detector was installed “on-site” to detect these markers at μ/m3 level in the atmosphere after pumping up the ambient air at close intervals. Quantitative results of the semi-continuous monitoring were strongly correlated with olfactive nuisances reported in the vicinity of the source. The chemical characterisation of odorous organic tracers and their “on-site” monitoring was shown to be a useful technique to detect malfunctions in the bio-gas collecting facilities and to prevent air pollution and olfactive nuisances in the environment.