IEEE Transactions on Instrumentation and Measurement · 1993 · 11 citations · 5 references
EngineeringElectrical EngineeringAnalog ElectronicsInstrumentationMicrowave EngineeringMicrowave MeasurementsReceiver SystemsAttenuation MeasurementAttenuatorsHigh-frequency AttenuationPrecision MeasurementSignal AttenuationVoltage
A homodyne system for measuring attenuation at 10 GHz and 30 MHz is constructed and evaluated. The radio frequency (RF) attenuation is converted to 1-kHz attenuation, and the 1-kHz attenuation is measured by an inductive voltage divider built in a 1-kHz receiver. A 1-kHz voltage ratio standard (VRS) is proposed to check the accuracy of the 1-kHz receiver for maintaining long-term accuracy. The differences between the VRS and the 1 kHz receiver are smaller than 0.0003 dB for 0-40 dB. A precision piston attenuator of 30 MHz agrees with the system within 0.0018 dB for 0-30 dB. The change in measured values at an interval of one year are less than 0.001 dB. The results demonstrate that this system can be expected to reliably maintain high accuracy.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">></ETX>
5
David H. Russell, Wilbur Larson · Proceedings of the IEEE · 1967 · 22 citations
Engineering, Electrical Engineering, Microwave Engineering +15