Concepedia

Publication | Closed Access

Modeling of ion processes in atmospheric pressure matrix‐assisted laser desorption/ionisation

16

Citations

8

References

2007

Year

Abstract

Atmospheric pressure matrix-assisted laser desorption/ionization (AP-MALDI) has proven a convenient and rapid method for ion production in the mass spectrometric analysis of biomolecules. This technique, like other atmospheric pressure ionization methods, suffers from ion loss during ion transmission from the atmosphere into the vacuum of the mass spectrometer. In this work, a simple model describing ion formation and ion motion towards the inlet capillary of the mass spectrometer is described. Both the gas flow and electric field near the MALDI plate were numerically calculated using the boundary element method (BEM). The ions were moving along with the gas flow and drifting in the electric field in accordance with their ion mobility properties. The ion signal dependence on an electric field strength obtained in the proposed model correlates well with experimental results.