A field-deployable, chemical ionization time-of-flight mass spectrometer

Timothy H. Bertram, Joel R. Kimmel, Timia A. Crisp, O. S. Ryder, R. L. N. Yatavelli, Joel A. Thornton, M. J. Cubison, M. Gonin, Douglas R. Worsnop

Atmospheric measurement techniques · 2011 · 342 citations · 30 references

DOIFull text

Open access

Concepts

TL;DR

The study develops a CI‑TOFMS for real‑time, high‑sensitivity measurement of atmospheric trace gases and demonstrates its applicability to diverse acids and reagent ions. The instrument features a high‑pressure interface with two RF‑only quadrupoles, efficient ion focusing through differential pumping, and a compact TOFMS, enabling real‑time measurement of trace gases via negative‑ion proton transfer with acetate. The CI‑TOFMS achieves >20 % ion‑duty cycle, >300 ions s⁻¹ pptv⁻¹ sensitivity, 10 Hz spectral acquisition, 5 % precision at 1 pptv, and a 4 pptv detection limit for formic acid. Abstract.

Abstract

Abstract. We constructed a new chemical ionization time-of-flight mass spectrometer (CI-TOFMS) that measures atmospheric trace gases in real time with high sensitivity. We apply the technique to the measurement of formic acid via negative-ion proton transfer, using acetate as the reagent ion. A novel high pressure interface, incorporating two RF-only quadrupoles is used to efficiently focus ions through four stages of differential pumping before analysis with a compact TOFMS. The high ion-duty cycle (>20 %) of the TOFMS combined with the efficient production and transmission of ions in the high pressure interface results in a highly sensitive (>300 ions s−1 pptv−1 formic acid) instrument capable of measuring and saving complete mass spectra at rates faster than 10 Hz. We demonstrate the efficient transfer and detection of both bare ions and ion-molecule clusters, and characterize the instrument during field measurements aboard the R/V Atlantis as part of the CalNex campaign during the spring of 2010. The in-field short-term precision is better than 5 % at 1 pptv (pL/L), for 1-s averages. The detection limit (3 σ, 1-s averages) of the current version of the CI-TOFMS, as applied to the in situ detection of formic acid, is limited by the magnitude and variability in the background determination and was determined to be 4 pptv. Application of the CI-TOFMS to the detection of other inorganic and organic acids, as well as the use of different reagent ion molecules (e.g. I−, CF3O−, CO3−) is promising, as we have demonstrated efficient transmission and detection of both bare ions and their associated ion-molecule clusters.

References

30