Radiation dose from a linear slit scanning X-ray machine with full-body imaging capabilities
30
Citations
16
References
2008
Year
X-ray FluorescenceRadiation ProtectionEffective DosesX-ray MachineLinear SlitPhoton-counting Computed TomographyInstrumentationRadiation ImagingRadiation OncologyCollimationRadiologyHealth SciencesRadiation DetectionMedical ImagingDigital X-ray EquipmentRadiographic ImagingDosimetryRadiation DoseMedicineX-ray Optic
Doses for a range of examinations and views using digital X-ray equipment with full-body linear slit scanning capabilities (Statscan) have been compared with those from other published studies. Entrance doses (free-in-air) were measured using a dosimeter, and effective doses were generated using a Monte Carlo simulator. Doses delivered by the linear slit scanning system were significantly lower than those from conventional X-ray equipment. Effective doses were between 9 and 75% of the United Nations Scientific Committee Report on the Effects of Ionising Radiation doses for standard examinations. This dose reduction can be explained by the properties of linear slit scanning technology, including low scatter, beam geometry, the use of a digital detector and the use of higher than usual tube voltages.