Publication | Open Access
Calibrating an interferometric laser frequency stabilization to megahertz precision
10
Citations
5
References
2012
Year
We report on a calibration procedure that enhances the precision of an interferometer based frequency stabilization by several orders of magnitude. For this purpose, the frequency deviations of the stabilization are measured precisely by means of a frequency comb. This allows us to implement several calibration steps that compensate different systematic errors. The resulting frequency deviation is shown to be less than 5.7 MHz (rms 1.6 MHz) in the whole wavelength interval 750-795 nm. Wide tuning of a stabilized laser at this exceptional precision is demonstrated.
| Year | Citations | |
|---|---|---|
Page 1