Simple, sensitive and rapid liquid chromatography/atmospheric pressure chemical ionization mass spectrometric method for the quantitation of Ranolazine in rat plasma

Yan Liang, Lin Xie, Xiaodong Liu, Weidong Chen, Tong Lu, Ye Xiong, Guangji Wang

Rapid Communications in Mass Spectrometry · 2006 · 16 citations · 4 references

Concepts

Abstract

A sensitive and specific method using liquid chromatography with atmospheric pressure chemical ionization mass spectrometry (LC/APCI-MS) has been developed and validated for the identification and quantification of Ranolazine in rat plasma. A simple liquid-liquid extraction procedure was followed by injection of the extracts onto a C18 column with isocratic elution and detection using a single quadrupole mass spectrometer in selected ion monitoring (SIM) mode. The method was tested using six different batches of blank plasma. Linearity was established for the concentration range 0.02-10.0 microg/mL, with a coefficient of determination (r) of 0.9998 and good back-calculated accuracy and precision. The intra- and inter-day precision (relative standard deviation (RSD) %) was lower than 10%, and accuracy ranged from 85 to 115%. The lower limit of quantification was reproducible at 0.01 microg/mL with 20 microL plasma. The proposed method enables the unambiguous identification and quantification of Ranolazine for pre-clinical and clinical studies.

References

4