IEEE Transactions on Applied Superconductivity · 1999 · 123 citations · 2 references
EngineeringBiomedical EngineeringElectrical EngineeringElectromagneticsMedical PhysicsSensorsMagnetic CharacterizationMagnetic SensorsMagnetic GradiometersMagnetic GradiometryMagnetoencephalographyMagnetometersWhite Noise
We have developed new systems for magnetoencephalography (MEG). The pick up coil is a coaxial type first-order gradiometer with 50-mm base line. The magnetic field resolution of the system is about 4 fT/Hz/sup 0.5/ or 0.8 (fT/cm)/Hz/sup 0.5/ in white noise region. The unique feature of the system is the gantry-free horizontal dewar, which is fabricated through what we call "ship-in-a-bottle approach". Less than 10-liter/day liquid helium consumption for the 100-liter capacity is realized. One of the merits of the horizontal dewar is that a small room is sufficient for installation because of the low height (0.89 m) of the dewar. Another merit is that the patient can be measured in lying position which is more relaxing compared to the conventional vertical type.
2
Integrated DC SQUID magnetometer with high dV/dB
D. Drung, R. Cantor, Mark Peters et al. · IEEE Transactions on Magnetics · 1991 · 84 citations
Elimination of flux-transformer crosstalk in multichannel SQUID magnetometers
H.J.M. ter Brake, F.H. Fleuren, J.A. Ulfrnan et al. · Cryogenics · 1986 · 55 citations · Full text