Three‐dimensional magnetic resonance imaging technique for myocardial‐delayed hyperenhancement: A comparison with the two‐dimensional technique

Kelly H. Zou, Mark Fruitman, H. Glenn Reynolds, Thomas Foo, Raymond Kwong, E. Kent Yucel

Journal of Magnetic Resonance Imaging · 2004 · 14 citations · 21 references

DOIFull text

Open access

Concepts

Abstract

PURPOSE: To compare two-dimensional and three-dimensional techniques in the detection of myocardial infarction (MI) and in the grading transmural extent (TE). MATERIALS AND METHODS: Twelve patients with clinically proven MI were examined using two-dimensional and three-dimensional techniques with cardiac-gated, breath-hold, T1-weighted gradient echo sequence with an inversion recovery pulse following gadopentetate dimeglumine (Gd-DTPA) at 0.2 mmol/kg. Contrast-to-noise, signal-to-noise, and signal intensity ratios (CNR, SNR, and SIR, respectively) were derived and compared for each technique. RESULTS: From two-dimensional to three-dimensional, statistical significant difference was found in the mean CNR (11.65 vs. 56.59; P = 0.002), SNR (18.03 vs. 76.90; P < 0.001), and SIR (3.6 vs. 6.36; P = 0.05). Intraobserver agreement (kappa) between two-dimensional and three-dimensional were R1 = 74% and R2 = 90%. Interobserver agreements between the readers were two-dimensional = 77% and three-dimensional = 79%. CONCLUSION: Mean CNR, SNR, and SIR are significantly increased in the three-dimensional technique compared to the conventional two-dimensional technique.

References