2002 · 118 citations · 4 references
EngineeringBiomedical EngineeringCardiologyBiomedical Signal ProcessingElectrodiagnostic MedicineElectrophysiologyCardiac Signal ProcessingElectrocardiogram AnalysisElectrocardiogram InterpretationElectrocardiographySignal VerificationVital Signs RecordingAutomated Ecg InterpretationEcg Morphology AnalysisEcg Quality AssessmentHolter Electrocardiogram AnalysisMisclassification CorrectionPrecisionT Wave AnalysisT Wave Detection
The authors evaluate a single-lead threshold based ECG wave boundaries detector with a QT database developed for validation purposes. They also identify its different sources of error distinguishing those that come from precision errors in boundary location from those that come from morphology misclassification. They obtain 71 % of records with correct morphology identification of T wave and variance in boundary location within manual referees variance. The remaining records analyzed correspond to signals with poor SNR at the T wave, or morphology discrepancies between algorithm and experts.