IEEE Transactions on Applied Superconductivity · 2004 · 11 citations · 6 references
MedicineElectrical EngineeringRadiologyDiagnostic ImagingElectromagneticsElectromagnetismMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsShielding DesignHigh-field Magnetic Resonance ImagingMagnetic Resonance Imaging ShieldingMagnetic ShieldingShielding
We have developed a new magnet for open high field (0.7 T) MRI systems. Since MRI systems are installed in hospital sites, it is important to reduce stray magnetic fields of the magnets. We studied passive, active and hybrid shielding methods and compared them. In order to study feasibility of active shielding, we considered higher order multipole stray fields. We present details of the calculation method and the results.
6
Superconducting open-configuration MR imaging system for image-guided therapy.
John F. Schenck, F.A. Jolesz, Peter Roemer et al. · Radiology · 1995 · 437 citations