Publication | Open Access
Patient safety concept for multichannel transmit coils
65
Citations
17
References
2007
Year
Our results demonstrate that current RF safety approaches are inadequate to deal with the new challenge of multichannel transmit coils. We propose a new concept based on a systematic investigation of the parameter space for RF phases and amplitudes. In this way the driving conditions generating the highest local SAR values per unit power can be identified and appropriately considered in the RF safety concept of a given MRI system.
| Year | Citations | |
|---|---|---|
Page 1
Page 1