Flow Cytometric Enumeration of Parasitemia in Cultures of<i>Plasmodium falciparum</i>Stained with SYBR Green I and CD235A

Jin Woo Jang, Ju Yeon Kim, Jung Hee Yoon, Soo Young Yoon, Chi Hyun Cho, Eun Taek Han, Seong Soo A. An, Chae Seung Lim

The Scientific World JOURNAL · 2014 · 18 citations · 21 references

DOIFull text

Open access

Concepts

Abstract

A flow cytometric (FACS) detection method for Plasmodium falciparum cultures (P. falciparum) was developed using SYBR Green I and CD235A and compared against the Giemsa stained microscopic examination. The cultured P. falciparum were spiked into red blood cells (RBCs) to yield parasitemia, ranging from 0.01% to 22.0%. FACS analysis demonstrated a clear separation between P. falciparum infected and uninfected RBCs. The measured percentage of parasitemia by FACS revealed higher precision (CV of 2.2-37.2%) with the sensitivity of 0.01% parasitemia than Giemsa stained microscopic examination (CV of 7.2-66.0%). High correlation of measured parasitaemia (r=0.98, P<0.05) was observed between FACS and Giemsa stained microscopic analyses. The higher levels of parasitaemia detection were observed in all ranges by FACS in comparison to Giemsa stained microscopic analysis. The currently reported FACS method using SYBR Green I and CD235A is potentially useful for measuring parasitemia in treating patients.

References

21