Sensors · 2015 · 153 citations · 36 references
MedicineNeurologyNeuroscienceMotor ControlMovement AnalysisMovement DisordersMovement KinematicsParkinson DiseaseParkinsonian DisordersPostural AssessmentTremor DisordersAction TremorInertial Measurement SensorsInertial SensorsParkinsonian TremorSpatiotemporal Time-frequency AnalysisTremor DiagnosticsTremor Severity AssessmentTremor Severity Rating
Quantitative assessment of parkinsonian tremor based on inertial sensors can provide reliable feedback on the effect of medication. In this regard, the features of parkinsonian tremor and its unique properties such as motor fluctuations and dyskinesia are taken into account. Least-square-estimation models are used to assess the severities of rest, postural, and action tremors. In addition, a time-frequency signal analysis algorithm for tremor state detection was also included in the tremor assessment method. This inertial sensor-based method was verified through comparison with an electromagnetic motion tracking system. Seven Parkinson's disease (PD) patients were tested using this tremor assessment system. The measured tremor amplitudes correlated well with the judgments of a neurologist (r = 0.98). The systematic analysis of sensor-based tremor quantification and the corresponding experiments could be of great help in monitoring the severity of parkinsonian tremor.
36
Clinically deployable Kinesia™ technology for automated tremor assessment
Joseph P. Giuffrida, David E. Riley, Brian N. Maddux et al. · Movement Disorders · 2009 · 247 citations