Archives of Ophthalmology · 1992 · 108 citations · 7 references
MedicineOrthopedic SurgerySurgeryImplant SurgeryOrthopaedic BiomaterialsOrthopedic BiomaterialsHydrogel BiomaterialsHydrogelsImplant ComplicationsImplant MaterialsHydrogel ImplantsHydrogel Swelling BehaviorLong-term ComplicationsLong-term Disease ComplicationsSoft Tissue Complications Around Implants
Seven cases in which long-term complications developed from swelling of the MAI hydrogel intrascleral buckling implant are reported herein. Micro-Fourier transform infrared spectroscopic analysis of two recovered implants demonstrated the occurrence of chemical changes leading to increased swelling. Clinical problems with the implants appeared 7 to 11 years after surgery, suggesting the need for periodic, long-term follow-up. It is possible that the present-day MIRAgel implant, which has the same chemical composition as the MAI implant, may require similar precautions.
7
HYDROGEL IMPLANT FOR SCLERAL BUCKLING
Felipe I. Tolentino, Manuela Roldán Pallarés, John M. Nassif et al. · Retina · 1985 · 45 citations
Medicine, Ophthalmology, Surgery +17
The MAI hydrophilic implant for scleral buckling: a review.
Patrick Ho, Ildefonso M. Chan, M.F. Refojo et al. · PubMed · 1984 · 34 citations
New hydrophilic implant for scleral buckling
M.F. Refojo · Annals of ophthalmology · 1980 · 32 citations
A HYDROPHILIC ACRYLATE IMPLANT FOR SCLERAL BUCKLING
Felipe I. Tolentino, Miguel F. Refojo, Charles L. Schepens · Retina · 1981 · 25 citations