Iranian Journal of radiation research/Iranian journal of radiation research · 2013 · 17 citations · 26 references
MedicineOncologyRadiation OncologyRadiation DosimetryRadiation MeasurementRadiation TransportExternal Beam RadiotherapyIntensity-modulated RadiotherapyPhoton DosimetryRadiation Dose MeasurementRadiation TherapySmall Field DosimetryBlockingExternal Beam Radiotherapy DosimetryGafchromic Film DosimetryIntensity-modulated Radiation Therapy DosimetryPartial Beam BlockingPhoton Small Field DosimetryStereotactic Radiosurgery DosimetryTherapeutic Dosimetry
Small photon ields are increasingly used in modern radiotherapy especially in intensity modulated radiation therapy (IMRT) and stereotactic radiosurgery (SRS) treatments. It is dif icult to measure beam pro iles, central axis depth doses and output factors in small ields accurately. That is related to the electron disequilibrium (1), partial blocking of the beam source (2, 3), steep dose gradients and perturbation due to different density of the detector and the medium (4, 5). Due to these M. Yarahmadi1,2, H.A. Nedaie2,3*, M. Allahverdi2, Kh. Asnaashari3, O.A. Sauer4
26
BEAM: A Monte Carlo code to simulate radiotherapy treatment units
D. W. O. Rogers, Bruce Faddegon, G Ding et al. · Medical Physics · 1995 · 1.4K citations
Precise radiochromic film dosimetry using a flat-bed document scanner
Slobodan Dević, Jan Seuntjens, E Sham et al. · Medical Physics · 2005 · 569 citations