Publication | Open Access
Evaluating the efficiency of isotope transmission for improved panel design and a comparison of the detection sensitivities of mass cytometer instruments
70
Citations
10
References
2015
Year
EngineeringMeasurementMicroscopyLanthanide DetectionIsotope TransmissionMicro Total Analysis SystemPathologyImmunophenotypingLanthanide CombinationsAnalytical InstrumentationCalibrationBioanalysisAnalytical ChemistryBiostatisticsClinical ChemistryInstrumentationLaboratory MedicineRadiation OncologyMass Cytometer InstrumentsChromatographyIsotope AnalysisRadiation DetectionHistopathologyDosimetryMass SpectrometryStable Isotope ProbingCell DetectionMedicineCytopathologyImproved Panel Design
The recent introduction of mass cytometry, a technique coupling a cell introduction system generating a stream of single cells with mass spectrometry, has greatly increased the number of parameters that can be measured per single cell. As with all new technology there is a need for dissemination of standardization and quality control procedures. Here, we characterize variations in sensitivity observed across the mass range of a mass cytometer, using different lanthanide tags. We observed a five-fold difference in lanthanide detection over the mass range and demonstrated that each instrument has its own sensitivity pattern. Therefore, the selection of lanthanide combinations is a key step in the establishment of a staining panel for mass cytometry-based experiments, particularly for multicenter studies. We propose the sensitivity pattern as the basis for panel design, instrument standardization and future implementation of normalization algorithms.