Errors in Dielectric Measurements Due to a Sample Insertion Hole in a Cavity

H. E. Bussey, A. J. Estin

IEEE Transactions on Microwave Theory and Techniques · 1960 · 59 citations · 5 references

Concepts

Abstract

The measurement of complex permittivities of isotropic media at microwave frequencies is performed with high precision by means of cylindrical cavity resonators. However, a hole in the cavity wall, for inserting the sample causes a frequency pulling of the resonator, which in turn introduces an error in the measured dielectric constant. These effects are measured, and with perturbation theory as a guide, correction factors are developed.

References

5