Automated Vulnerability Testing Identifies Patients With Inadequate Defibrillation Safety Margin

10

Citations

15

References

2012

Year

Abstract

An automated system identified all patients who failed conventional safety margin testing, while inducing only 19% of patients. Although limited by sample size, this study suggests the feasibility of automated implant testing that substantially reduces the need for VF induction in patients receiving implantable cardioverter-defibrillators.