A strain gauge plethysmograph device with a DC bridge supplied from a constant current source is described. The theoretical advantage of using a constant current source is discussed. An electrical calibration unit facilitates the clinical use of the plethysmograph. The electrical calibration method is compared with a mechanical one. The frequency response of the plethysmo graph, with the strain gauges mounted in working position is studied.
13
The measurement of volume changes in human limbs
R. J. Whitney · The Journal of Physiology · 1953 · 1.5K citations · Full text
Medicine, Anatomy, Physiology +14
The relation of pulsatile pressure to flow in arteries
D. A. McDonald · The Journal of Physiology · 1955 · 218 citations