Clinical Chemistry · 1990 · 79 citations · 25 references
PharmacyPharmaceuticsPharmacologyAnalytical ChemistryAnalytical ToxicologyChromatographic AnalysisAnalytical ChromatographyChromatographyGas ChromatographyLiquid ChromatographyBlood Drug AnalysisDrug Detection TechnologiesHigh-performance Liquid ChromatographyIllicit Drug DetectionInternal Standard AnalysisReversed-phase ChromatographyReversed-phase Liquid ChromatographyStimulant Detection
We describe a procedure for measuring concentrations of cocaine in plasma by reversed-phase high-performance liquid chromatography, with ion-pairing. The procedure involves solvent extraction followed by back-extraction with dilute acid. The n-propyl ester of benzoylecgonine is used as an internal standard. An evaporation step is not required, and concentrations as low as 5 micrograms/L can be quantified. Plasma concentrations of cocaine determined by high-performance liquid chromatography correlated well with those determined by a previously reported gas-chromatographic procedure with nitrogen detection.
25