IEEE Transactions on Nuclear Science · 2005 · 10 citations · 12 references
EngineeringRadiologyDiagnostic ImagingRadiation DetectionSensorsDetector TechnologyImage SensorsRadiation Imaging DetectorsRadiation Imaging SystemsRadiographic ImagingX-ray DetectorsCmos Image SensorsDigital Radiography DetectorsFiber-optic SensorsImage Sensor DesignRadiographic EquipmentRadiographic Image QualityX-ray Imaging PhysicsX-ray Imaging Systems
As a continuation of our digital X-ray imaging sensor R&D, we have developed a cost-effective, portable, digital radiographic system based on a CMOS image sensor coupled with a fiber optic plate (FOP) and selected conventional scintillators. The imaging system consists of a commercially available CMOS image sensor of 48 /spl mu/m /spl times/ 48 /spl mu/m pixel size and 49.2 mm /spl times/ 49.3 mm active area, a FOP bundled with several millions of glass fibers of about 6 /spl mu/m in diameter and 3 mm in thickness, phosphor screens such as Min-R or Lanex series, a readout IC board, a GUI software, and a battery-operated X-ray generator (20-60 kV/sub p/; up to 1 mA). Here the FOP was incorporated into the imaging system to reduce the performance degradation of the CMOS sensor module caused by irradiation and also to improve image quality. In this paper, we described each imaging component of the fully-integrated portable digital radiographic system in detail, and also presented its performance analysis with experimental measurements and acquired X-ray images in terms of system response with exposure, contrast-to-noise ratio (CNR), modulation transfer function (MTF), noise power spectrum (NPS), and detective quantum efficiency (DQE).
12
DQE(<i>f</i>) of four generations of computed radiography acquisition devices
James T. Dobbins, David L. Ergun, Lois Rutz et al. · Medical Physics · 1995 · 307 citations