Concepedia

Publication | Closed Access

Comparison of apically extruded debris after large apical preparations by full‐sequence rotary and single‐file reciprocating systems

100

Citations

31

References

2015

Year

Abstract

Abstract Aim To evaluate the amount of apically extruded debris produced by ProTaper Universal ( PTU ), ProTaper Next ( PTN ), WaveOne ( WO ) and Reciproc (R) systems after large apical preparations. Methodology Sixty mandibular premolars with a single canal were selected and randomly assigned into 4 groups ( n = 15) according to the system used for root canal preparation: PTU , PTN , WO and R groups. Canal preparations were performed up to size 40 in each group. Distilled water was used as an irrigant, and the apically extruded debris from each tooth was collected in pre‐weighted glass vials and dried. The average weight of debris was assessed using a microbalance, and the data were analysed statistically using one‐way analysis of variance and the post hoc Tukey multiple comparison test (α = 0.05). Results The PTU system was associated with significantly more debris than the other systems ( P < 0.05). No significant differences were found between PTN , WO and R systems ( P > 0.05). Conclusions All systems were associated with apical debris extrusion when canals were prepared to a large apical size. The PTU system was associated with more debris extrusion.

References

YearCitations

Page 1