Ultrasound in Medicine & Biology · 2014 · 16 citations · 27 references
Natural SciencesBioengineeringBiomedical EngineeringAcousticsUltrasonographyUltrasoundAcoustic MaterialsAcoustic MeasurementPhysical AcousticsUltrasonicsUltrasound InstrumentationUltrasound PhysicsAcoustic PropertiesCarrageenan BiomaterialsMaterial Acoustic PropertiesSpeed Of Sound MeasurementUltrasonic AcousticsUltrasonic ImagingUltrasonic Materials CharacterizationUltrasound-based Tissue Characterization
The acoustic properties of a robust tissue-mimicking material based on konjac–carrageenan at ultrasound frequencies in the range 5–60 MHz are described. Acoustic properties were characterized using two methods: a broadband reflection substitution technique using a commercially available preclinical ultrasound scanner (Vevo 770, FUJIFILM VisualSonics, Toronto, ON, Canada), and a dedicated high-frequency ultrasound facility developed at the National Physical Laboratory (NPL, Teddington, UK), which employed a broadband through-transmission substitution technique. The mean speed of sound across the measured frequencies was found to be 1551.7 ± 12.7 and 1547.7 ± 3.3 m s21, respectively. The attenuation exhibited a non-linear dependence on frequency, f (MHz), in the form of a polynomial function: 0.009787f2 1 0.2671f and 0.01024f2 1 0.3639f, respectively. The characterization of this tissue-mimicking material will provide reference data for designing phantoms for preclinical systems, which may, in certain applications such as flow phantoms, require a physically more robust tissuemimicking material than is currently available.
27
V. A. Del Grosso, C. W. Mader · The Journal of the Acoustical Society of America · 1972 · 890 citations
A wall-less vessel phantom for Doppler ultrasound studies
Daniel W. Rickey, Paul A. Picot, D.A. Christopher et al. · Ultrasound in Medicine & Biology · 1995 · 236 citations