A 128-Electrode Three Dimensional Electrical Impedance Tomography System

Guizhi Xu, Renping Wang, Shuai Zhang, Shuo Yang, Gusphyl Justin, Mingui Sun, Weili Yan

Conference proceedings · 2007 · 19 citations · 4 references

Concepts

Abstract

Electrical impedance tomography (EIT) is a new functional imaging technique. This paper presents the development of a new electrical impedance tomography system with 128 electrodes for impedance change detection and 3D imaging of the human thorax. The system consists of several modules, including multi-frequency current source, driving, measuring, data acquisition, and controlling and signal processing modules. A high speed digital signal processor (DSP) is used as the controller. The 64 driving electrodes and 64 measuring electrodes are positioned uniformly in four planes around the surface of a cylindrical phantom filled with a saline solution and objects of varying conductivities. The performance has been tested, and these preliminary experiments demonstrate the feasibility this system.

References

4