IEEE Transactions on Biomedical Engineering · 1999 · 60 citations · 20 references
EngineeringBiomedical EngineeringElectrical EngineeringBiomedical Signal ProcessingElectrophysiologyExercise PhysiologyCardiac Signal ProcessingDigital Signal ProcessingElectrocardiogram AnalysisElectrocardiographyExercise ElectrocardiographySignal Processing12-Lead ElectrocardiographyBaseline Wander CorrectionBaseline Wander RemovalChannelsEcg Signal ReconstructionSingular Value Analysis
An orthogonalization method to eliminate unwanted signal components in standard 12-lead exercise electrocardiograms (ECG's) is presented in this work. A singular-value-decomposition-based algorithm is proposed to decompose the signal into two time-orthogonal subspaces; one containing the ECG and the other containing artifacts like baseline wander and electromyogram. The method makes use of redundancy in 12-lead ECG. The same method is also tested for reconstruction of a completely lost channel. The online implementation of the method is given. It is observed that the first two decomposed channels with highest energy are sufficient to reconstruct the ST-segment and J-point. The dimension of the signal space, on the other hand, does not exceed three. Data from 23 patients, with duration ranging from 9 to 21 min, are used.
20
A Real-Time QRS Detection Algorithm
Jiapu Pan, W.J. Tompkins · IEEE Transactions on Biomedical Engineering · 1985 · 7.7K citations