Removal of phase artifacts from fMRI data using a Stockwell transform filter improves brain activity detection

Bradley G. Goodyear, Hongmei Zhu, Robert A. Brown, J. Ross Mitchell

Magnetic Resonance in Medicine · 2003 · 55 citations · 15 references

DOIFull text

Open access

Concepts

TL;DR

The Stockwell transform (ST) is a mathematical operation that yields the frequency content at each time point within a time‑varying signal. The study presents an automated method to remove fMRI artifacts caused by motion outside the imaging field of view. The method applies 1D Fourier transforms to raw fMRI data to obtain phase profiles, then uses the ST on the phase‑magnitude time series to locate and excise artifact frequencies by replacing them with median values before applying an inverse ST to reconstruct the signal. ST‑filtering significantly improves fMRI brain activity detection by removing artifacts overlapping regions of interest while preserving other frequency information. Published in Magn Reson Med 51:16–21 (2004); © 2003 Wiley‑Liss, Inc.

Abstract

Abstract A novel and automated technique is described for removing fMRI image artifacts resulting from motion outside of the imaging field of view. The technique is based on the Stockwell transform (ST), a mathematical operation that provides the frequency content at each time point within a time‐varying signal. Using this technique, 1D Fourier transforms (FTs) are performed on raw image data to obtain phase profiles. The time series of phase magnitude for each and every point in the phase profile is then subjected to the ST to obtain a time‐frequency spectrum. The temporal location of an artifact is identified based on the magnitude of a frequency component relative to the median magnitude of that frequency's occurrence over all time points. After each artifact frequency is removed by replacing its magnitude with the median magnitude, an inverse ST is applied to regain the MR signal. Brain activity detection within fMRI datasets is improved by significantly reducing image artifacts that overlap anatomical regions of interest. The major advantage of ST‐filtering is that artifact frequencies may be removed within a narrow time‐window, while preserving the frequency information at all other time points. Magn Reson Med 51:16–21, 2004. © 2003 Wiley‐Liss, Inc.

References

15