Current Eye Research · 2008 · 41 citations · 47 references
Life SciencesImmunologyImmunopathologyImmunoregulationInflammationBioanalytical Method ValidationCytokine BiologyCytokine RegulationFlow CytometryBead-based AssaysCytokine BioassaysInflammatory Cytokine Response AssaysLuminex Bead AssaysTear Cytokine AnalysisTear CytokinesTear Film BiomarkersTear Inflammatory Biomarkers
PURPOSE: To compare instruments and reagents used for cytometric bead-based assay (CBA) of cytokines in human tears. METHODS: Two CBA instruments and reagent kits from five vendors were compared in parallel. Vendor protocols were modified to accommodate tear volumes. Precision, recovery, dilution linearity, and stability were evaluated. RESULTS: Luminextrade mark and Bio-Radtrade mark proved optimal for tear CBA. Good correlation for dilution linearity was observed (r > 0.80, p < 0.05) for 1-4-mu l sample volumes. Samples were stable for 30 min at 25 degrees C. CONCLUSION: Many tear cytokines can be reliably measured by CBA using an optimized protocol. Concentrations remain stable during periods consistent with collection.