Background effects on Faraday collectors in gas‐source mass spectrometry and implications for clumped isotope measurements

173

Citations

13

References

2013

Year

Abstract

Background determination with gas flowing into the source of the mass spectrometer is necessary for accurate clumped isotope measurements of CO(2). Background corrections can be performed accurately if the slit width of the m/z 44 Faraday cup significantly exceeds that of the one for m/z 47, using a correlation between m/z 44 signal intensity and the corresponding minimum in m/z 47 background. We propose two new correction schemes that reduce the time-consuming measurement of gases of different bulk isotopic compositions. These findings may also be relevant for the measurement of other rare isotopologues by mass spectrometry.