Journal of Physics Conference Series · 2010 · 10 citations · 10 references
MedicineRadiation OncologyRadiologyRadiation DosimetryRadiation MeasurementTreatment PlanningBrachytherapy DosimetryBrachytherapy PhysicsExternal Beam DosimetryRadiation TherapyThree-dimensional Dose DistributionTreatment Planning Dose CalculationBrachytherapy Dose CalculationBrachytherapy Dose DistributionBrachytherapy Dosimetry VerificationDose Distribution CalculationsImaging-based DosimetryLow-dose DosimetrySource DosimetryThree-dimensional Dose Calculation
PURPOSE: The feasibility of using the PRESAGE/Optical-CT system for 3D dosimetry verification around a brachytherapy source is investigated. METHOD AND MATERIALS: Brachytherapy dose distributions were obtained by irradiation of cylindrical PRESAGE volumes 6cm in diameter by 8cm height with a GammaMed 12i Ir-192 HDR unit (Varian Medical Systems). A narrow channel on the central axis was created by setting a steel catheter in the Presage during manufacture, enabling measurements close to the source (~3mm). RESULTS: Comparison of dose line profiles shows good agreement between PRESAGE and verified calculated dose calculation, in both high and low dose regions. CONCLUSION: The PRESAGE/Optical-CT shows good potential in verification of 3D dose distributions around brachytherapy sources.
10
CERR: A computational environment for radiotherapy research
Joseph O. Deasy, Angel I. Blanco, V Clark · Medical Physics · 2003 · 876 citations
A comprehensive evaluation of the PRESAGE/optical‐CT 3D dosimetry system
H Sakhalkar, J Adamovics, Geoffrey S. Ibbott et al. · Medical Physics · 2008 · 111 citations · Full text