Determination of plutonium concentration and its isotopic ratio in environmental materials by ICP-MS after separation using and extraction chromatography

116

Citations

13

References

1999

Year

Abstract

An analysis method for 239 Pu and 240 Pu in environmental samples (including four certified reference materials) by ICP-MS was studied. Two types of chromatographic resin, Dowex 1X8 and TEVA, were examined for their applicability to the separation of Pu from the matrix elements. Sufficient decontamination factors (10 4 -10 5 ) for many matrix elements including U, which interferes with the detection of mass 239, were obtained with both resins. The detection limit of Pu by ICP-MS was about 0.02 pg ml –1 (0.05 mBq ml –1 for 239 Pu; 0.17 mBq ml –1 for 240 Pu) in the sample solution or 0.1 pg in the sample. Analytical results of 239+240 Pu in certified reference materials (IAEA-135, -SOIL-6, -368 and -134) indicated that the accuracy of the method was satisfactory. Data on the 240 Pu/ 239 Pu atom ratios in these reference materials, which are scarce in the literature, were also obtained, i.e., 0.211 for IAEA-135, 0.191 for IAEA-SOIL-6, 0.043 for IAEA-368 and 0.200 for IAEA-134. Compared with alpha-spectrometry, the ICP-MS method has significant advantages in terms of its simple analytical procedures, prompt measurement time and capability of determining the 240 Pu/ 239 Pu ratio.