Publication | Open Access
Microembolic signals and strategy to prevent gas embolism during extracorporeal membrane oxygenation
62
Citations
9
References
2010
Year
We established that the microembolic load depends on gas embolism from the central venous lines and on the level of blood flow assistance.The gas microemboli that enter in the blood circulation and in the extracorporeal circuits are not removed by the membrane oxygenator filter.Maximum care is required in drugs and fluid infusion of this kind of patients as a possible source of microemboli. This harmful phenomenon may be overcome adding an air filter device to the intravenous catheters.