Development and Validation of High-Performance Liquid Chromatography–Diode Array Detector Method for the Determination of Tramadol in Human Saliva
12
Citations
39
References
2012
Year
Human SalivaMedicineHealthy VolunteersBioanalysisDrug TestAnalytical ChemistryDrug MonitoringSaliva ConcentrationsClinical ChemistryLiquid ChromatographyPharmacologyPharmacokineticsChromatographyAnesthesiologyDrug Analysis
A high-performance liquid chromatography–diode array detector method (HPLC-DAD) was developed and validated for the determination of tramadol in human saliva. Samples of saliva were prepared utilizing liquid–liquid extraction with hexane-ethyl acetate (4:1, v/v). Propranolol was used as an internal standard (IS). The linear range and average recovery of tramadol were 0.25–4.00 μg/mL and 94.70%, respectively. The intra- and inter-assay precisions at low, intermediate and high concentrations were within 1.40–9.79 % and 0.77–3.98%, respectively. Intra- and inter-assay accuracies were within 98.50–107.13% and 100.44–103.89%, respectively. The proposed method was successfully applied to determine tramadol in saliva concentrations in three healthy volunteers for 24 hours after administration of 100 mg oral doses of tramadol.