IEEE Transactions on Nuclear Science · 1997 · 94 citations · 7 references
MedicineNuclear MedicineRadiologyRadiation DosimetryRadiation MeasurementTreatment PlanningOncologic Nuclear ImagingPositron Emission TomographyRadiation Dose MeasurementRadiation Imaging SystemsRadiation TherapyTreatment Planning Dose CalculationDose Distribution CalculationsDose Distribution MappingFragmentationHeavy Ion RadiationIon Beam DosimetryNuclear FragmentationPet QuantificationRadiation Dose Estimation
A method for the in-situ control of the heavy-ion tumour therapy by means of positron emission tomography is introduced. This method is founded on the measurement of the dynamic spatial distributions of /spl beta//sup +/-emitters generated by nuclear fragmentation during the irradiation and their relation to the dose. In order to study this relationship and to derive the dose distribution from the measured /spl beta//sup +/-activity distribution, a framework of model calculations is used. Results of phantom experiments with /sup 12/C ion beams are presented, demonstrating good agreement between experiment and calculation as well as the possibilities and limits of the PET-technique for treatment plan verification and beam monitoring.
7
The investigation of different cameras for in-beam PET imaging
Jörg Pawelke, L. Byars, W. Enghardt et al. · Physics in Medicine and Biology · 1996 · 83 citations